Fluke PM3370B user manual RS-232 Cable Configurations, Rts, Cts, Dtr, Dsr, Dcd, Gnd

Models: PM3370B

1 297
Download 297 pages 53.38 Kb
Page 285
Image 285

RS-232 CABLE CONFIGURATIONS

N - 1

Appendix N

RS-232 Cable configurations

This appendix supplies additional information about the RS-232 cable configurations between the oscilloscope and a connected device. The oscilloscope and most of the devices are Data Terminal Equipment (DTE) configurated.

Communication lines:

TxD

=

Transmitted Data

RxD

=

Received Data

RTS

=

Request To Send

CTS

=

Clear To Send

DTR

=

Data Terminal Ready

DSR

= Data Set Ready (= Modem Ready)

DCD

=

Data Carrier Detect

S.GND

=

Signal ground

F.GND

=

Safety ground

A.Cable to printer/plotter/controller with software handshake parameters: "3-wire/7-wire" = 3-wire

"XON-XOFF on off" = on

OSCILLOSCOPE

DEVICE

(DTE)

(DTE)

SIGNAL

9 PIN

9 PIN

25 PIN

 

(female)

 

 

TxD

3

3

2

RxD

2

2

3

RfR

7

7

4

CTS

8

8

5

DTR

4

4

20

DSR

6

6

6

DCD

1

1

8

S.GND

5

5

7

F.GND

CASE

CASE

1

 

 

 

ST6930

ATTENTION:

The maximum Baud rate for the PM8277 and PM8278 plotters is

 

4800 baud. For a higher Baud rate you must use a hardware

 

handshake configuration.

NOTE:

The LQ1500 printer can have a 6-pole DIN connector with the

 

following connections:

 

 

TxD on pin 1

DTR on pin 2

 

RxD on pin 3

S.GND on pin 5

Page 285
Image 285
Fluke PM3370B user manual RS-232 Cable Configurations, Rts, Cts, Dtr, Dsr, Dcd, Gnd