6 - 46THE CPL PROTOCOL

TRIGGER ACQUISITION

TA

Purpose:

To perform a software trigger.

This command causes an acquisition or sweep to be started. It is the software equivalent of a normal trigger pulse. In the single shot mode the AT (Arm Trigger) command is sent to arm the triggering first.

Command:

"TA"

Response:

CR

acknowledge

Example:

CR

100 PRINT #1,"TA" :’Send Trigger Acquisition

110 GOSUB 1000 :’Sync on acknowledge