FUNCTION REFERENCE

5 - 59

REMOTE CONTROL RS-232

Description:

The oscilloscope is equipped with an RS-232 Interface as standard. This can be used for remote control or for setting the readout using an external controller or PC.

The language used is called CPL (Compact Programming Language) and is described in Chapter 6. CPL is a small set of very powerful commands for full remote control of all oscilloscope functions. The male 9-pin connector and its connections are shown in the figure below.

For correct functioning Communication parameters must be adjusted. This is done in the menu UTILITY RS-232 SETUP. This menu is reached via the key sequence ’UTILITY >> RS-232 SETUP >> ......’. Possible Selections are:

-Baud rate. To be selected with the TRACK control.

-Number of DATAbits and PARITY. Combinations are

dataBITS:

7

7

8

8

8

PARITY:

Odd

Even

Odd

Even

No

There is always one stopbit.

 

 

 

-Hardware handshake is selected with ’3-wire/7-wire’. In the 7 wire position the hardware handshake signals DSR/DTR and CTS are active.

-Software handshake is selected with ’XON-XOFF on off’.

Operation of the front panel key STATUS LOCAL passes the control of the oscilloscope from the interface (remote) to the front panel keys (local).