HOW TO USE THE INSTRUMENT

4 - 47

CURSORS

READOUT

-Press the softkey next to ’READOUT’.

CURSORS

READOUT

The ’CURSOR READOUT’ selection menu for horizontal and vertical measurements is now displayed. You can make the following selections from this menu: ΔT, 1/ΔT, ΔT-ratio, T-trg and phase for time measurements.

ΔV, V1, V2, ΔV-ratio for voltage measurements.

ch1: ΔT= 460s ΔV= 600mV

ΔT 1/ΔT ΔT-ratio ph T-trg

ΔV

V1&V2 ΔV-ratio

RETURN

ST6732

9312

ΔT is the time difference between the time cursors. The readout is in seconds (s) or fractions of seconds.

1/ΔT results in the inverted value of the time difference, and is displayed in Hz. When the time cursors are set exactly one period apart this represents the frequency of the signal.

ΔT-ratioallows for a relative measurement of two cursor readings. It is displayed as a percentage.

Phase is used to measure a phase difference between two signals, or to determine the phase of a detail within one signal.

T-trggives the timedifference between the cursors and the trigger point.

ΔV is the voltage difference between the voltage cursors (readout: ΔV=...V). V1, V2 displays the absolute voltages for each voltage cursor in relation to ground (readout: V1=Vreference, V2=Vdelta).

ΔV-ratioallows for a relative measurement of two cursor readings. It is displayed as a percentage.

Page 92
Image 92
Fluke PM3370B Cursors Readout, Press the softkey next to ’READOUT’, ΔV, V1, V2, ΔV-ratio for voltage measurements