5 - 40

FUNCTION REFERENCE

AUTO

Toggle key to switch the AUTO RANGE function on/off

RANGE

AMPL

Toggle key to switch between two vertical input sensitivities of the External Trigger input channel.

INPUT COUPLING

Description:

The characteristics of the oscilloscope’s inputs can be selected for each channel. A channel can be switched on/off with the toggle key ON. Display of the External Trigger input signal is switched on/off with the toggle key TRIG VIEW. CH1 is always switched on when all other channels are switched off.

In the dc coupled mode the complete signal including dc components and extremely low frequencies (<10 Hz) are displayed. In the ac coupled mode, dc components are suppressed. This results in a display of the ac components (10 Hz).

Ground coupling (GND) interrupts the input signal. The position of the trace is at the 0 volt level. A continuous ground level indication for each channel can also be activated. Refer to the description of the UTILITY>>SCREEN&SOUND MENU function.

The selections are made with the toggle keys AC/DC/GND for the channel inputs, or AC/DC for the External Trigger input.

The type of coupling is given in the readout area using the symbols (ac), = (dc) and (GND).

0dB

3dB

0dB

AC-COUPLING 3dB

DC-COUPLING

10Hz

FULL BANDWITH

DC

FULL BANDWITH

 

 

 

FREQ.

 

 

 

FREQ.

 

 

 

 

 

MAT4219

Effect of ac/dc input coupling

Page 160
Image 160
Fluke PM3370B Input Coupling, Toggle key to switch the Auto Range function on/off, Effect of ac/dc input coupling