5 - 12

FUNCTION REFERENCE

AUTOSET USERPROG

Description:

A number of instrument settings after AUTOSET can be customized to your special needs when the standard default settings do not fit on your application.

The selections are reached via the key sequence ’UTILITY >> AUTOSET >> userprog’. This gives access to VERT, TRIG and PROBE submenus where the selections are made. The following table shows possible selections. Unaffect means that existing selections are not overruled by AUTOSET.

Description

Standard

Userprog

Menu

 

AUTOSET

alternatives

UTILITY>>AUTOSET

 

 

 

>>userprog>>.

 

 

 

 

Channel selection

volt/div, on/off

unaffect

VERT

Input coupling

ac

dc, unaffect

VERT

Input Impedance

1 MΩ

50Ω, unaffect 1)

VERT

Bandwidth limiter

off

on, unaffect

VERT

Triggering

edge, ac,

unaffect

TRIG

 

level-pp on,

 

 

 

auto, LEVEL

 

 

 

MTB center

 

 

Probe

manual probe

unaffect

PROBE

 

selection set

manual probe

 

 

to 1:1

selections

 

 

 

 

 

1)200 MHz models only

Page 132