Fluke PM3370B user manual Cursors Readout Time, Measured

Models: PM3370B

1 297
Download 297 pages 53.38 Kb
Page 142
Image 142
MAT4210
Δ
-POSITIONΔ CURSOR ON NEGATIVE SLOPE OF PULSE
ch ... :ΔT=25%

5 - 22

FUNCTION REFERENCE

CURSORS READOUT

Description:

The cursors offer a wide variety of voltage and time readouts. For comparison of signal details the ratio mode is very suitable. When in analog mode time or volt cursors have been selected, only the relevant readouts are displayed in the READOUT menu. In digital mode all readouts are displayed at the same time.

The various readout selections for time, volt and ’both’ are reached via the key sequence ’CURSORS >> READOUT’.

CURSORS READOUT

TIME

Three time interval readouts can be selected:

-ΔT: Gives the time between the cursors.

-1/ΔT: This results is a frequency readout. The readout is correct when the distance between the cursors equals one signal period.

-ΔT-ratio: The readout is a percentage

that can be reset by using the

ch ... :ΔV=100%

 

 

 

 

ΔT=100 %’ softkey. This mode can be

 

 

 

used for time comparisons. The figure

 

 

 

shows an example of a duty cycle

 

 

 

measurement. First the cursors are

 

 

 

positioned at the period and the readout

 

 

 

is set to 100%. Then the pulse width is

 

 

-POSITION CURSORS

measured.

TRACK

Δ

ON 1 SIGNAL PERIOD

 

 

 

-PRESS ΔT=100%

-Phase (ph): The readout is a number of

degrees that can be reset using the ’ΔT=360 °’ softkey.

-T-trg (time to trigger): The readout gives the time between the trigger point and each time cursors.

Page 142
Image 142
Fluke PM3370B user manual Cursors Readout Time, Measured