4 - 32HOW TO USE THE INSTRUMENT

ACQUIRE

AVERAGE

The average function averages the input data over a number of successive acquisitions. The average function is used to reduce the influence of random noise in the input signal. There is no loss of bandwidth when the average function is activated, but the signal must be repetitive. The number of samples over which the average is calculated can be selected by the user.

-Use the probe to connect Ch1 to the Probe Adjust output.

-Press AUTOSET.

-Connect the second probe to channel 2, but do not connect it to a signal!

-Press the CH1+CH2 key to add Ch1 and Ch2 together.

-Press the mV key of the CH2 AMPL keys so that the noise picked up by the probe tip of the second probe appears on the CH1+CH2 trace.

-Press the AVERAGE key.

For maximum ease of use, the scope offers direct access to the important average feature.

The text ’Average = 8’ appears in the bottom of the screen to indicate that the function is active. The default value is 8. The indication changes into ’AVG’. Observe that the noise on the trace is reduced.

-To change the Average constant, press the ACQUIRE key.

-Turn the TRACK control clockwise.

The Average factor increases when the TRACK control is turned clockwise. Observe that this reduces noise on the trace even more.

 

TRACK

Average=2

Average=256

AVG

AVG

ST6481

-Turn the TRACK control counterclockwise so that the Average factor is 8 again.

Page 77
Image 77