FUNCTION REFERENCE

5 - 25

Every time the signal crosses this level the event counter is incremented by one. When the event counter reaches the selected delay value, the scope triggers and a new signal acquisition is started. The number of events to be counted before the acquisition starts is selected by using the Δ control in the EVENT DELAY sub menu.

Event delay and trigger position can be used in combination to offer powerful delay capabilities.

CH1

CH2

ACQUISITION:

CH1: SLOPE

TRIGGER

POSITION

CH2: EVENT, COUNT=8

 

TRIGGER POSITION: .... ms

 

ST6768

Key sequence:

Control to adjust of the trigger position.

Toggle softkey to switch the Event mode on and off. When the Event mode is turned on, further menu selections are displayed

Control to change the number of events. Readout in the menu is called ‘COUNT’.

Toggle softkey to select the channel on which the events must be counted.

Control to set the events level. Readout in the menu is called ‘LEVEL’.

Toggle softkey to select the positive or negative slope.

Page 145
Image 145