6 - 12THE CPL PROTOCOL

GO to REMOTE

GR

Purpose:

Puts the oscilloscope in the Remote State. In the Remote State none of the oscilloscope functions are accessible via the front panel buttons and knobs.

Going back to the Local State is achieved by sending the Go to Local (GL) command or by pressing the STATUS/LOCAL key on the frontpanel (Refer also to "Local Lockout" and "Go to Local").

Command:

"GR"

CR

Response:

acknowledge

CR

Example:

100

PRINT

#1,"GR"

:’Send

command

110

INPUT

#1,ACK

:’Sync

on acknowledge

Local, Remote, Remote+Local Lockout States

power-on

Local State

"Go to Remote"

"Go to Local" "Reset Instrument"

<STATUS/LOCAL button>

Remote State

"Local Lockout"

"Go to Local" "Reset Instrument"

Remote with

Local Lockout

State

Page 229
Image 229
Fluke PM3370B user manual GO to Remote, 12THE CPL Protocol