Fluke Rev. 4 Continuity Testing, Frequency Measurement, 701/702, Users Manual, Tone, Resistance

Models: Rev. 4

1 118
Download 118 pages 56.12 Kb
Page 100
Image 100
Continuity Testing

701/702

Users Manual

Continuity Testing

TONE

Continuous Tone

May or may not get tone

No tone

RESISTANCE

<25 Ω

25 to 400 Ω >400 Ω

Frequency Measurement

RANGES

1 Hz to 109.99 Hz

110 Hz to 1099.9 Hz

1.100 kHz to 10.999 kHz

11 kHz to 50 kHz

Amplitude range for frequency measurement: <1 kHz: 110 mV to 300V

1 kHz to 30 kHz: 0.5 to 30V

>30 kHz: 1V to 30V Input Impedance: 5 MΩ

 

MODEL 701 AND 702

 

COUNTS

1 YEAR

 

 

 

2 YEAR

5

 

 

5

5

 

 

5

5

 

 

5

5

 

 

5

For frequency measurement less than 109.99 Hz, specification applies for signals with a slew rate greater than 5 volt/millisecond.

90

Page 100
Image 100
Fluke Rev. 4 Continuity Testing, Frequency Measurement, 701/702, Users Manual, Tone, Resistance, Ranges, MODEL 701 AND