Fluke Rev. 4 Loop Power Supply, Documenting Process Calibrator, Specifications, MODELS 702 AND

Models: Rev. 4

1 118
Download 118 pages 56.12 Kb
Page 109
Image 109
Sensor inaccuracies not included

Documenting Process Calibrator

Specifications

Temperature, Resistance Temperature Detectors (cont)

100Ω Pt(3916)

-200 to -190

-190 to 0

0 to 630

200Ω, 500Ω, 1000Ω Pt(385)

0.3

0.3

0.5

0.4

0.4

0.8

0.3

0.1

0.2

0.4

0.2

0.4

-200 to 0

0 to 400

400 to 630

0.3

0.5

0.8

0.5

0.8

1.0

0.1

0.2

0.4

0.2

0.4

0.5

Sensor inaccuracies not included

Resolution: 0.1°C

Temperature Coefficient: 0.02°C/°C in the ranges -10 to 18°C and 28 to 50°C

Maximum Input Voltage: 30V

For two and three-wire RTD measurements, add 0.4°C to the specifications.

Loop Power Supply

MODELS 702 AND 701

SETTING

24 Volt

28 Volt

Short circuit protected Maximum Current: 22 mA Maximum Input Voltage: 30V dc

1YEAR 5% 5%

2YEAR 5% 5%

99

Page 109
Image 109
Fluke Rev. 4 Loop Power Supply, Documenting Process Calibrator, Specifications, Maximum Input Voltage, MODELS 702 AND