9.Press the As Found softkey. The display changes as follows:

10.Enter values for 0% and 100% of 4.0 mA and 20.0 mA, respectively. Set Tolerance to 0.5% of span. (Use other values if necessary for your application.)

User Entered Value is for a value measured by some other device, such as the reading reported by the control room. Leave this set to Disabled because you are only interested in values measured at the transmitter output.

NOTE

If you set User-Entered Value to Enabled, you will be prompted to enter a value at each test point in the measured units. Errors will be calculated and displayed based on the user-entered value, not the measured value.

11.Use the arrow keys to move the cursor down to enter 0% and 100% values for SOURCE temperature. Our example uses 100°C and 300°C.

12.The Test Strategy is the number of test points and which test points are performed rising and falling in percent of scale. Our example uses five points (0%, 25%, 50%, 75%, and 100%), rising only. Rising is indicated by the up arrow on the display. Change to another test strategy by pressing eon this line. A list of strategies appears from which to choose. Select one, then press Done. When you finish selecting the calibration parameters, the display appears as follows: