Using Measure Mode

Documenting Process Calibrator

Using Measure Mode

5.The ID string is shown at the bottom of the boxed area. To erase a character, press the Back Space softkey. To erase the whole string, press c.

6.Press u, d, L, or Rto select a character, then press e. Use the numeric keypad if you want to enter a number.

7.Repeat step 6 until you are satisfied with the ID string appearing in the window.

8.Press the Done softkey.

9.Press the Done softkey or sto exit s mode.

Using Measure Mode

NOTE

To achieve best measurement noise and accuracy performance, do not use the battery eliminator, and tie all three common jacks together.

The operating mode (i.e., MEASURE, SOURCE) is shown in a reverse-video bar on the display. If the calibrator is not in MEASURE mode, press Muntil MEASURE is shown. You must be in MEASURE mode to change any of the MEASURE parameters.

Measurement Ranges

The calibrator normally changes to the appropriate measurement range automatically. The lower right side of the display shows either “Range” or “Auto Range” depending on the range status. Auto Range switch points are shown in the Specifications at the end of this manual. When you press the Rkey, the range is locked. Press it again to cycle to and lock on the next higher range. Press it and hold it for two seconds to revert to Auto Range.