Documenting Process Calibrator

Specifications

DC Current Measurement

RANGE

RESOLUTION

MODEL 702

% OF READING +% OF FULL SCALE

MODEL 701

% OF READING +% OF FULL SCALE

30mA

110mA

1μA

10μA

1 YEAR

0.025% + 0.025%

0.05% + 0.05%

2 YEAR

0.025% + 0.03%

0.07% + 0.07%

1 YEAR

0.025% +0.025%

0.05% + 0.05%

2 YEAR

0.025% + 0.03%

0.1% + 0.07%

Temperature Coefficient: (0.001% of rdg. + 0.002% f.s.)/°C in the ranges -10 to 18°C and 28 to 50°C

Common Mode Error: 0.01% f.s./(Common Mode Volt)

Maximum Input Voltage: 30V dc

Resistance Measurement

RANGE

RESOLUTION

MODEL 702

% OF READING + Ohms

1 YEAR

2 YEAR

MODEL 701

% OF READING + Ohms

1 YEAR

2 YEAR

11Ω

110Ω

1.1 kΩ

11 kΩ

0.001 Ω

0.01Ω

0.1Ω

1 Ω

0.05% + 0.05

0.05% + 0.05

0.05% + 0.5

0.1% + 10

0.075% + 0.05

0.075% + 0.05

0.075% + 0.5

0.1% + 10

0.1% + 0.05

0.1% + 0.05

0.1% + 0.5

0.1% + 10

0.15% + 0.05

0.15% + 0.05

0.15% + 0.5

0.15% + 10

Temperature Coefficient: (0.01% f.s. + 2 mΩ) /°C in the ranges -10 to 18°C and 28 to 50°C

Common Mode Error: 0.005% f.s/(Common Mode Volt)

Maximum Input Voltage: 30V dc