Fluke Rev. 4 user manual Frequency Sourcing, Frequency Ranges

Models: Rev. 4

1 118
Download 118 pages 56.12 Kb
Page 104
Image 104

701/702

Users Manual

Frequency Sourcing

FREQUENCY RANGES

2 Hz to 109.99 Hz

110 Hz to 1099 Hz

1.1 kHz to 10.9 kHz

11 kHz to 50 kHz

Waveform: Squarewave, 50% duty cycle

Amplitude: 0 to 10V pk

Amplitude Accuracy:

2 to 1099 Hz: 3% of output + 0.05 V

1.1to 10.9 kHz: 10% of output + 0.05 V 11 to 50 kHz: 30% of output + 0.05 V

Maximum Input Voltage: 30V dc

 

MODELS 702 AND 701

 

COUNTS

1 YEAR

 

 

 

2 YEAR

1

 

 

1

1

 

 

1

1

 

 

1

2

 

 

2

94

Page 104