Documenting Process Calibrator

Specifications

DC Voltage Output

RANGE RESOLUTIONS

MODEL 702

% OUTPUT +% OF FULL SCALE

MODEL 701

% OUTPUT + % OF FULL SCALE

110 mV

1 μV

1.1 V

10 μV

11 V

100 μV

1 YEAR

0.01% + 0.005%

0.01% + 0.005%

0.01% + 0.005%

2 YEAR

0.015% + 0.005%

0.015% + 0.005%

0.015% + 0.005%

1 YEAR

0.015% + 0.005%

0.015% + 0.005%

0.015% + 0.005%

2 YEAR

0.02% + 0.005%

0.02% + 0.005%

0.02% + 0.005%

Temperature Coefficient: (0.001% of output + 0.001% of f.s.)/°C in the ranges -10 to 18°C and 28 to 50°C

Maximum Output Current: 10 mA

Loading: (0.001% f.s. + 1 μV)/ mA

Common Mode Error: 0.008% f.s/(Common Mode Volt)

Maximum Input Voltage: 30V dc

91