Fluke Rev. 4 DC Voltage Measurement, Specifications, Documenting Process Calibrator, Range, Year

Models: Rev. 4

1 118
Download 118 pages 56.12 Kb
Page 97
Image 97
Specifications

Documenting Process Calibrator

Specifications

DC Voltage Measurement

RANGE

110mV

1.1 V

RESOLUTION

1μV

10μV

MODEL 702

% OF READING +% OF FULL SCALE

1 YEAR
2 YEAR

0.025% + 0.015%

0.05% + 0.015%

0.025% + 0.005%

0.05% + 0.005%

MODEL 701

% OF READING +% OF FULL SCALE

1 YEAR

2 YEAR

0.05% + 0.015%

0.1% + 0.015%

0.05% + 0.005%

0.1% + 0.005%

11 V

100 μV

0.025% + 0.005%

0.05% + 0.005%

0.05% + 0.005%

0.1% + 0.005%

110 V

1 mV

0.05% + 0.005%

0.1% + 0.005%

0.05% + 0.005%

0.1% + 0.005%

300 V

10 mV

0.05% + 0.005%

0.1% + 0.005%

0.05% + 0.005%

0.1% + 0.005%

Temperature Coefficient: (0.001% of rdg. + 0.0015% f.s.)/°C in the ranges -10 to 18°C and 28 to 50°C

Input Impedance: 5 MΩ

Common Mode Error: 0.008% f.s./(Common Mode Volt)

Maximum Input Voltage: 300V rms