AC Voltage Measurement

701/702

Users Manual

AC Voltage Measurement

FREQUENCY RANGE

MODEL 701 AND 702

(% OF READING + NUMBER OF COUNTS)

1 YEAR

2 YEAR

20 to 40 Hz

40 to 500 Hz

500 Hz to 1 kHz

1 kHz to 5 kHz

2% + 10

0.5% + 5

2% + 10

10% + 20

2% + 10

0.5% + 5

2% + 10

10% + 20

Specifications apply for 10% to 100% of voltage range Ranges : 1.1000 V, 11.000 V, 110.00 V, 300.0 V rms

Resolution: 11.000 counts in all ranges except 300V; 3,000 counts on 300V range.

Input Impedance: 5 MΩ and <100 pF

Temperature Coefficient: 10% of spec/°C in the ranges -10 to 18°C and 28 to 50°C

Input Coupling: ac
Maximum Input Voltage: 300 V rms

For frequencies >1 kHz, the minimum voltage is 0.5 V