Fluke Rev. 4 user manual Documenting Process Calibrator, Using Source Mode

Models: Rev. 4

1 118
Download 118 pages 56.12 Kb
Page 67
Image 67
12.To start ramping, press the Start Ramp softkey.

Documenting Process Calibrator

Using Source Mode

8.Fill in the parameters as prompted. Enter the Start Value, End Value, and Ramp Time.

9.If you want to automatically stop ramping if a trip condition is detected, set the Trip Detect to Enabled, and select Voltage or Continuity as the trip function.

10.Press the Done softkey. Note the RAMP annunciator next to SOURCE at the top of the display.

11.Select a low-to-high ramp or a high-to-low ramp with the Ramp Up/Down softkey.

12.To start ramping, press the Start Ramp softkey.

13.Ramping continues until a trip is detected (if enabled), the ramp time expires, or you press the Stop Ramp softkey.

57