Sourcing in Percent of Scale

701/702

Users Manual

Sourcing in Percent of Scale

To adjust the source value as a percent of scale as shown below, proceed as follows:

1.If necessary, press Mfor SOURCE mode.

2.Select a source function (m, v, q, t, or p) as previously described.

3.Press the Scale softkey.

4.Use the numeric keypad to enter the 0% of scale value.

5.Press e.

6.Use the numeric keypad to enter the 100% of scale value.

7.Press e.

8.Press the Done softkey.

Percent of scale remains in effect until you press the Scale softkey again.

Stepping and Ramping the Output Value

Two special features are available for adjusting the value of source functions except pressure (sourcing pressure requires that you use an external pressure source):

Stepping the output with the uand dkeys.

Ramping the output with optional continuity or V trip detect.

Stepping the Output

The step feature allows you to select a step size in engineering units (mV, V, mA, °C, etc.) or % of scale. Stepping the output in % of scale is useful for quickly jumping between 0% and 100% (set step size = 100%) or 0-50-100% (set step size = 50%). Stepping