Fluke Rev. 4 user manual Automatically Saving Battery Life, Using the Optional Battery Eliminator

Models: Rev. 4

1 118
Download 118 pages 56.12 Kb
Page 33
Image 33
Automatically Saving Battery Life

Documenting Process Calibrator

Setting Up the Calibrator

Automatically Saving Battery Life

An optional Auto Battery Save feature turns the calibrator off after a selected idle time. The default setting for Auto Battery Save is Off. When Auto Battery Save is On, the Esymbol shows in the upper right corner of the display. The setting is preserved after you turn off the power. Auto Battery Save works the same on battery power and when using the battery eliminator. Turn on the Auto Battery Save feature as follows:

1.Press s.

2.Press d to highlight Off following Auto Battery Save.

3.Press eor the Choices softkey.

4.Press u to highlight On, then press e.

5.To accept the timeout period shown on the display, you can finish here. Press Done to exit smode and do not go on to step 6.

6.To change the timeout period, press d to highlight the timeout period following Battery Save Timeout.

7.Press eor the Choices softkey.

8.Enter your choice of timeout period in minutes (accepted range: 1 to 120 minutes).

9.Press the Done softkey.

10.Press the Done softkey or sto exit s mode.

Using the Optional Battery Eliminator

CAUTION

To avoid damage to the calibrator, use only Fluke Model BE9005 Series Battery Eliminator, available from your distributor.

Where ac power is available, you can use the optional Fluke Model BE9005 Battery Eliminator to conserve battery power. When the battery eliminator is used, the battery is internally disconnected, and can be removed from the calibrator. The calibrator operates the same as it does under battery power, however, the specifications for the 701/702 only apply under battery operation. The battery eliminator does not charge the battery.

23

Page 33
Image 33
Fluke Rev. 4 Automatically Saving Battery Life, Using the Optional Battery Eliminator, Documenting Process Calibrator