Adjusting the Transmitter

701/702

Users Manual

value is shown in the top left of the measure window.

15.The calibrator moves on to the remaining set of points. For temperature and electrical parameter calibration, the points are done automatically. If you are sourcing pressure, the calibrator pauses at each step for you to adjust the pressure source. When the tests are complete, an error summary table, such as the following is displayed.

failures were outside the ±0.5% tolerance that we selected.

17.Either press the Done softkey to save the data, or the Abort softkey to delete the data and start over. You have to go back to As Found to restart the calibration. You can see the saved data entry and recall the table for later viewing through the Review Memory softkey during normal operation. With Model 702 you can upload this data to a PC using PMLink software. See the PMLink Manual for instructions.

16.In the results summary test, failures are highlighted. An adjustment is required in this example because three tests show failures. The

Adjusting the Transmitter

Proceed as follows to make the calibration adjustments to the transmitter. (Always refer to the transmitter manufacturer’s instructions to locate the adjustment controls and connection points for your transmitter).

1.Press the Done softkey while viewing the results summary.

2.Press the Adjust softkey. The calibrator sources 0% of span (100°C in this example) and displays the following softkeys:

66