Quick Guide to Applications

Documenting Process Calibrator

Quick Guide to Applications

the list of tasks (procedures) downloaded from PMLink. Tasks are 702 calibrator configurations, saved with a procedure name, for example the type and manufacturer of a specific transmitter. A procedure configures the calibrator for transmitter calibration with all the calibration parameters (SOURCE/MEASURE functions, 0% and 100% levels, test strategy) predefined. See the PMLink manual for instructions on developing and downloading tasks.

Quick Guide to Applications

The following figures show test lead connections and which calibrator function to use for many different applications.

Clearing Memory

The Clear Memory softkey, shown after you press More Choices softkey three times, clears all the memory:

Saved results

Min Max data

Log data sets (Model 702 only)

A confirmation message appears so that you do not inadvertently erase the memory.