Documenting Process Calibrator
Specifications
Resistance Sourcing
RANGES
RESOLUTION
MODEL 702
% OUTPUT + OHMS
1 YEAR | 2 YEAR |
MODEL 701
% OUTPUT + OHMS
1 YEAR | 2 YEAR |
11.000Ω
110.00Ω
1.1000 kΩ
11.000 kΩ
1 mΩ
10 mΩ
100 mΩ
1Ω
0.01% + 0.02
0.01% + 0.04
0.02% + 0. 5
0.03% + 5
0.02% + 0.02
0.02% + 0.04
0.03% + 0.5
0.04% + 5
0.02% + 0.02
0.02% + 0.04
0.05% + 0.5
0.1% + 5
0.03% + 0.02
0.03% + 0.04
0.1% + 0.5
0.15% + 5
Temperature Coefficient: (0.01% of f.s.)/°C in the ranges
Maximum/Minimum Current through Source Resistance: 11Ω Range: 3 mA dc max, 0.1 mA dc min
110Ω Range: 3 mA dc max, 0.1 mA dc min
1.1kΩ Range: 3 mA dc max, 0.01 mA dc min 11 kΩ Range: 1 mA dc max, 0.01 mA dc min
Common Mode Error: 0.008% f.s/(Common Mode Volt)
Maximum Input Voltage: 30V dc
93