Graco 332564B important safety instructions Enter Measured Volume of Solvent

Models: 332564B

1 104
Download 104 pages 58.79 Kb
Page 79
Image 79

Setup Mode Screens

Calibrate Screen 3

Calibrate Screen 3 initiates a calibration of an accessory solvent meter. During the test, the Volume Verification screen will appear.

The meter and lines must be primed with solvent before doing the calibration. See Solvent Meter Calibration, page 86 for complete instructions.

To initiate the calibration, press the Volume Check button.

The screen displays the volume dispensed. Enter

the amount of solvent dispensed in the Measured

 

Figure 46 Enter Measured Volume of Solvent

Volume field, or press

to end the test.

After the Measured Volume is entered, the Accept

Calibration window will appear. Press to accept

the calibration. Press to cancel the calibration and retain the previous K-factor.

Press and hold the Reset button for 1-2 seconds to reset the volume counter.

Figure 47 Accept Calibration

Figure 45 Calibrate Screen 3

332564B

79

Page 79
Image 79
Graco 332564B important safety instructions Enter Measured Volume of Solvent