Calibration Screens

Calibrate Screen 1

Calibrate Screen 1 initiates a pump pressure check (stall test) for the selected pump. During the test, the Stall Test screen will appear.

The pump and lines must be primed with color or catalyst before doing the stall test. See System Screen 2, page 68 to set test parameters. See Pump Pressure Check, page 84 for complete test instructions.

To initiate the test, press the Pressure Check button for the desired pump. The pump will build pressure in the line to a minimum of the Stall Test Pressure. The pump will then move to the center stroke position and stall test the upstroke, followed by the downstroke.

NOTE: The Last Passed log can only be reset by successfully completing the test.

The screen displays the number of days since the last stall test was passed for each pump.

Setup Mode Screens

Figure 41 Calibrate Screen 1

Figure 42 Stall Test Screen