Chapter 12. CALIBRATION

Built-in clock adjustment

￿￿￿

 

￿￿￿￿ ￿￿￿

￿

 

￿

￿

 

 

 

Figure 12-4.

 

 

 

Example:

 

A: 2 digits of calendar year

1993 ……………………93

 

B: Month

April ……………………04

 

C: Day

5th day …………………05

 

D: Hour (24 hour system)

3PM ……………………15

 

E: Minute

6 minutes ………………06

 

F: Second

Not changeable

 

NOTE

1., , , , keys are used to change the values, and ENTER key is used to store the values.

2.Clock is still progressing the operation even after the calibration step is moved to this built-in clock adjustment item. After entering the changing mode by ENTER key, the clock stops.

3.After pressing ENTER key, the clock starts from 0 second.

4.Writing to EEPROM is not necessary.

PV calibration

Scroll , , , , keys to show (1011) on upper display, then press ENTER key.

Input CH No. select

Press SETUP key until the PROG/SEG display shows (01-01).

Input the channel No. by , , , , keys ( ENTER key not required).

Table 12-8.

 

 

￿

0 0 0 0 0

0 0 0 0 1

 

 

Gain No. select

Press SETUP key until the PROG/SEG display shows (01-02). Connect calibration device (See Section 12-3 Set Up”).

Input the gain number (See Table 12-9 and Table 12-10) by scrolling , , , , keys the upper display ( ENTER key not required).

12-11