4-3.How To Operate Maintenance Mode

"Re-printing Printed Page (PREVIOUS PRINT)

In this device, as the data that has already been printed is stored in the inner hard disk, re-printing can be made later. In this case, it is possible to change the print preferences such as the LUT applied for printing, print density, and contrast, etc.

Follow the procedure below to re-print.

NOTE

The print data in the device is stored for each diagnostic device. The page number is assigned to each print data in order in which it is received from the diagnostic device and printed out. (The page No. is printed on the upper part of the film as the film management information.)

Film management information

Page No.

 

 

1 0 0 1 P 300 20 32768 662 115301771170003 1,0R01

CH1:REGIUS_350 PAGE:4

 

 

(1)Touch [PREVIOUS PRINT] on the maintenance menu screen.

The PREVIOUS PRINT screen (diagnostic device select screen) will appear.

The names of the connected diagnostic devices are listed on the diagnostic device select screen.

(2)Touch the name of the diagnostic device from which the print data is received to print.

The selected diagnostic device name will be highlighted. When 10 or more diagnostic devices are connected, touch [$] / [%] to scroll the list of the diagnostic device names.

(3)Touch on the upper part of the screen.

The print page set screen will appear.

20 DRYPRO MODEL 752 Operation Manual Ver.0.11 2002.6

Page 32
Image 32
Konica Minolta 752 Re-printing Printed Page Previous Print, Touch Previous Print on the maintenance menu screen