(5)Touch  on the upper part of the screen.

4-3.How To Operate Maintenance Mode

(4)Select the diagnostic device name you wish to apply as the exposure conditions of test printing.

The selected diagnostic device name will be highlighted. When 10 or more diagnostic devices are connected, touch [$] / [%] to scroll the list of the diagnostic device names.

(5)Touch Manual background on the upper part of the screen.

The LUT select screen will appear.

(6)Touch a LUT name you wish to apply for printing.

The selected user LUT name will be highlighted.

(7)Touch Manual background on the upper part of the screen.

The print preference set screen (2/3) will appear.

(8)Touch the key on the right of the setting item, then touch [$] / [%] to set the following items.

DENSITY

: Adjust the density of an image.

 

Selectable from -7 to 7.

 

Setting a negative value will make the den-

 

sity lighter, and a positive value, darker.

CONTRAST

: Adjust the contrast of an image.

 

Selectable from -7 to 7.

 

Setting a negative value will make the con-

 

trast lower, and a positive value, higher.

SMOOTH TYPE : Sets the smoothing type of an image.

 

 

SMOOTH TYPE

Settings

 

 

 

OFF

 

Does not perform the magnification

 

 

interpolation.

 

 

 

1

 

Performs the Biliner interpolation.

 

 

 

2-6

 

Performs the Cubic Spline interpo-

 

 

lation.

 

 

Increasing the value will smooth an

 

 

image greater.

 

 

 

DRYPRO MODEL 752 Operation Manual Ver.0.11 2002.6 27