Konica Minolta 752 operation manual Calibrating Density CALIBRATION

Models: 752

1 87
Download 87 pages 20.91 Kb
Page 36
Image 36
"Calibrating Density (CALIBRATION)

4-3.How To Operate Maintenance Mode

"Calibrating Density (CALIBRATION)

The device has the densitometer in the cooling/discharge section, which allows you to control the density automatically.

If there is a problem with the density, print the calibration sheet from this screen to calibrate the density.

Calibration

(1)Touch [CALIBRATION] on the maintenance menu screen.

The message for selecting the tray will appear.

(2)Touch the key of the tray for which you wish to per- form density correction.

The print data in the calibration sheet will be queued, and the maintenance menu screen will re-appear.

If you wish to perform density correction for the other tray, repeat from Step (1).

(3)Touch [EXIT] to end the maintenance mode.

When the normal screen re-appears, the film in the selected tray is transported and the calibration sheet is printed out. Wait until printing is complete. After printing is end, the density cor- rection is complete.

IMPORTANT

The calibration sheet is used as the sampling data (the data before the density correction) for the automatic density correction.

When you check and control the density, use the SMPTE pattern which is printed out by TEST PRINT on the maintenance menu. (Refer to page 26.)

24 DRYPRO MODEL 752 Operation Manual Ver.0.11 2002.6

Page 36