4-3.How To Operate Maintenance Mode
"Creating user LUT
Although this device has the LUT that is suitable for the diagnostic device at shipping, you also can create user's own LUT by inputting the LUT characteristics. Follow the procedure below to create the user LUT.
(1)Touch [USER LUT] on the maintenance menu screen.
The user LUT create screen (USER LUT) listing the name of the connected diagnostic devices will appear.
(2)Touch a name of the diagnostic device on which the user LUT is applied.
The selected diagnostic device name will be highlighted. When 10 or more diagnostic devices are connected, touch [$] / [%] to scroll the list of the diagnostic device names.
(3)Touch
on the upper part of the screen.
The user LUT select screen will appear.
(4)Touch a LUT name you wish to apply on the diag- nostic device.
The selected LUT name will be highlighted.
(5)Touch [DATA] on the lower part of the screen.
The LUT data screen (next page) will appear.
By touching [LUT LIBRARY] instead of [DATA], the user LUT can by set using the LUT in the library stored in this device. (Refer to "Setting user LUT" on page 42.)
DRYPRO MODEL 752 Operation Manual Ver.0.11 2002.6 43