Konica Minolta 752 operation manual Automatic calibration

Models: 752

1 87
Download 87 pages 20.91 Kb
Page 37
Image 37

4-3.How To Operate Maintenance Mode

Automatic calibration

The DRYPRO has the automatic density calibration function that perform the density calibration automatically after a lapse of the specified time or specified number of printing. (Instead of performing automatic calibration, you can make [CALIBRA- TION] appear on the screen to show you that the calibration needs performing depending on the settings at installation.)

Or the automatic density calibration can be performed when the film is loaded.

CAUTION

Even after a lapse of the specified period or specified number of printing, the automatic calibration is not performed if printing data is not queued. The automatic calibration is per- formed when the printing data is queued after a lapse of the specified period or specified number of printing.

The automatic calibration is performed if the printing data is not queued even when the film is loaded. The automatic calibration is per- formed when the printing data is queued on the next time.

DRYPRO MODEL 752 Operation Manual Ver.0.11 2002.6 25

Page 37
Image 37
Konica Minolta 752 operation manual Automatic calibration