5-12 Chapter 5: Calibration

Procedure 1: Increasing Load Test

This test checks the scale’s accuracy by incrementally adding weight to the center of the Top Plate. Use weights that correspond to the NCR 7870 weight feature. If this test fails, recalibrate scale.

Instructions

1. Select the appropriate weight feature from the table below.

Step

Weight Feature

Add

Remove

Display

 

 

 

Weight

Weight

 

 

 

 

 

 

 

 

 

1

9.995 kg

0.1 kg

 

0.1 ±

0.00 kg

 

13.995 kg

0.1 kg

 

0.1 ±

0.00 kg

 

30.0 lb.

0.2 lb.

 

0.2 ±

0.00 lb.

2

9.995 kg

2.5 kg

0.01 kg

2.5 ±

0.00 kg

 

13.995 kg

2.5 kg

0.01 kg

2.5 ±

0.00 kg

 

30.0 lb.

5.0 lb.

0.2 lb.

5.0 ±

0.00 lb.

3

9.995 kg

2.5 kg

 

5.0 ±

0.005 kg

 

13.995 kg

4.5 kg

 

7.0 ±

0.005 kg

 

30.0 lb.

5.0 lb.

 

10.0 ±

0.01 lb.

4

9.995 kg

2.5 kg

 

7.5 ±

0.005 kg

 

13.995 kg

3.0 kg

 

10.0 ±

0.005 kg

 

30.0 lb.

10.0 lb.

 

20.0 ±

0.01 lb.

5

9.995 kg

2.495 kg

 

9.995 ±

0.005 kg

 

13.995 kg

3.995 kg

 

13.995 ±

0.005 kg

 

30.0 lb.

10.0 lb.

 

30.0 ±

0.01 lb.

2.Follow steps shown in table for the weight feature of your scale.

3.Check readout after completing each step and verify the measured value falls within the acceptable limits.

4.If a measured value does not fall within limits, recalibrate scale. If all measured values are acceptable, the scale has passed the Increasing Load Test. Go to Procedure 2: Over-Capacity Test.

Note: Do NOT remove any weights from the Top Plate.