ALLU

ALL Use

Classification:

Format, Image and Sequence memory management

Command Syntax:

ALLU

Query Syntax:

None

Description:

The ALLU command first checks the current contents of the

 

format buffer for errors. If no errors are found, it reconfigures

 

the signal generating hardware in accordance with the contents.

 

Next, the current test image is re-rendered using the latest system

 

and format parameter data.

Other Required Cmds:

This command updates the generator after using the FMTL, IMGL

 

and SEQL commands to load new files from memory. This

 

command also can be used to see the results of work when

 

using commands to edit formats or custom images.

Example:

F M T L v g a _ m 3 / / L o a d a f o r m a t f r o m m e m o r y

/ / t o b u f f e r

A L L U / /

U p d a t e h a r d w a r e t o c u r r e n t

/ /

b u f f e r c o n t e n t s

ASBG

Analog Sync on Blue Gating

Classification:

Format parameter setting

Command Syntax:

ASBG

Limits:

<mode>

 

0 = OFF

 

1 = ON

Query Syntax:

ASBG?

Query Response:

<mode>

Description:

The ASBG command enables and disables adding composite sync

 

to the blue analog video outputs when analog sync is selected

 

(see SSST command) and an analog video signal is being generated

 

(see AVST command). The ASBG? query returns the current

 

setting of ASBG.

 

See the ASSG command description for information on

 

simultaneously controlling red, green and blue sync gating.

Other Required Cmds: The FMTU command instructs the generator to use the new setting. The ALLU command updates hardware to the new setting and redraws the test image.

Example: A S B G 1 / / E n a b l e c o m p s y n c o n b l u e i n / / b u f f e r

F M T U / / U p d a t e h a r d w a r e t o c u r r e n t / / b u f f e r c o n t e n t s

6-87

Page 213
Image 213
Quantum Data 801GC, 801GF, 801GX manual Allu, ALL Use, Asbg, Analog Sync on Blue Gating