VSIZ

Vertical SIZe

Classification:

Format parameter setting

Command Syntax: VSIZ <physical size>

Limits:

<physical size> = positive value (floating point accepted)

Query Syntax:

VSIZ?

Query Response:

<physical size>

Description:

The VSIZ command establishes the vertical physical size of the

 

image on the display. Units expected (or returned) vary according

 

to the last mode set with USIZ command. The VSIZ command

 

is context sensitive and must appear between begin and end

 

commands: FMTB and FMTE. The VSIZ? query returns the current

 

setting of VSIZ.

 

 

 

 

Note – Make sure that the USIZ parameter is properly set before

 

using the VSIZ command. Changing the USIZ setting after

 

entering VSIZ will convert the size to match the new unit of

 

measure.

 

Other Required Cmds:

The units of measure must be properly set by USIZ before entering

 

VSIZ. The ALLU command updates hardware to the new setting

 

and redraws the test image, taking the new size into account.

Example:

F M T B / / B e g i n e d i t i n g s e s s i o n

 

/ / O n e o r m o r e f o r m a t e d i t i n g

 

/ / c o m m a n d s

 

/ / . . .

 

U S I Z 1 / / S e l e c t i n c h e s a s u n i t o f / / m e a s u r e i n b u f f e r

H S I Z 1 0 . 4 / / S e t w i d t h t o 1 0 . 4 i n / / b u f f e r

V S I Z 7 . 8 / / S e t h e i g h t t o 7 . 8 i n b u f f e r A L L U / / T e s t t h e n e w s e t t i n g s

/ / O n e o r m o r e f o r m a t e d i t i n g / / c o m m a n d s

/ / . . .

F M T E / / E n d e d i t i n g s e s s i o n

6-172

Chapter 6: Programming

Model 801GC, 801GF & 801GX¥Rev. A

Page 298
Image 298
Quantum Data 801GC, 801GF, 801GX manual Vertical SIZe, 172, Vsiz?