Quantum Data 801GC, 801GF, 801GX Vertical physical size too small, Pedestal swing out of range

Models: 801GC, 801GF, 801GX

1 356
Download 356 pages 47.77 Kb
Page 324
Image 324

2706 Vertical physical size too small

The physical vertical size must be greater than zero.

Change the vertical physical size to a positive, nonzero value.

2714 Pedestal swing out of range

The analog video blanking pedestal cannot be less than 0 IRE (%) or more than 100 IRE (%).

Change the value of Pedestal level to be greater than or equal to 0 and less than or equal to 100 IRE (%).

2715 Gamma correction out of range

The firmware does not support gamma correction factors that are less than 0.1 or greater than 10.0.

Change the value of Gamma to be greater than or equal to 0.1 and less than or equal to 10.0.

2716 Analog video swing out of range

The hardware does not support peak-to-peak analog video swings that are less than 0 or greater than 1.000 volts. The swing does not include any composite sync levels that may be added.

Change the Video swing to be greater than or equal to 0 and less than or equal to 1.000 volts.

2717 Sync swing out of range

The hardware does not support peak-to-peak analog composite sync swings that are less than 0 or greater than 0.400 volts.

Change the Sync swing to be greater than or equal to 0 and less than or equal to 0.400 volts.

2719 Video swing calibration out of range

One or more of the analog video swing calibration factors are out of range. These parameters are not actually part of a Format that is stored in memory. They are system level parameters that affect how all analog Formats are read from memory and into the hardware. The parameter settings are maintained while the 801GP is turned off.

Re-initialize your generator’s memory.

Contact Quantum Data for further assistance.

8-16

Chapter 8: Error Messages

Model 801GC, 801GF & 801GX¥Rev. A

Page 324
Image 324
Quantum Data 801GC, 801GF, 801GX Vertical physical size too small, Pedestal swing out of range, Sync swing out of range