HRAT

Horizontal RATe

Classification:

Format parameter setting

Command Syntax: HRAT <frequency in Hz>

Limits:

<frequency in Hz> (floating point accepted)

 

typical min = 1000

 

typical max = 130000

Query Syntax:

HRAT?

Query Response:

<frequency in Hz> (floating point returned)

Description:

The HRAT command sets the line frequency. Pixel rate is equal

 

to the product: HTOT times HRAT. Frame rate is equal to the

 

quotient: HRAT divided by VTOT. Field rate is equal to the

 

product: SCAN times the frame rate. The HRAT? query returns

 

the current horizontal frequency setting.

Other Required Cmds: The FMTU command instructs the generator to use the new setting. The ALLU command updates hardware to the new setting and redraws the test image.

E x a m p l e s : H R A T 3 2 7 6 8 / / S e t 3 2 . 7 6 8 k H z r a t e i n / / b u f f e r

F M T U / / U p d a t e h a r d w a r e t o c u r r e n t

 

/ / b u f f e r c o n t e n t s

 

or

 

H R A T 3 2 . 7 6 8 E 3 / / S e t a 3 2 . 7 6 8 k H z r a t e

 

F M T U / / U p d a t e h a r d w a r e t o c u r r e n t

 

/ / b u f f e r c o n t e n t s

HRES

Horizontal RESolution

Classification: Format parameter setting

Command Syntax: HRES <pixels>

Limits: <pixels> min = 16

max depends on VRES and model of generator

Query Syntax:

HRES?

Query Response:

<pixels>

Description:

The HRES command establishes the number of active pixels per

 

line. The HRES? query returns the current setting of HRES.

Other Required Cmds:

The ALLU command updates hardware to the new setting and

 

redraws the test image.

Example:

H R E S 4 8 0 / / S e t 4 8 0 a c t i v e p i x e l s l i n e

/ / i n b u f f e r

A L L U / / C o n f i g u r e h a r d w a r e a n d r e d r a w / / i m a g e