4002 No sequence memory

This occurs during the SEQN command. It happens when there’s an insufficient amount of managed memory for the buffer request.

Save all used edit buffers and issue FMTN, IMGN, and/or DIRN. This clears extra memory used by these buffers.

4003 Sequence memory full

There is not enough free memory to save the sequence that is in the edit buffer.

Reduce the number of steps in the sequence being edited.

Delete one or more sequences currently stored in memory. You will need to abandon your current edited sequence to do this.

4005 Seq ed running

This occurs during SEQB, SEQN, or SEQE commands. It happens when you have used the Sequence Editor and then try to use the SEQB or SEQN command.

Exit the generator's GUI sequence editor before editing a sequence using commands via the communications port.

4006 No sequence to save

This occurs during SEQA or SEQS commands. It happens when you try to save the contents of the sequence editing register and none is initialized.

You must use SEQN before SEQS or SEQA.

4007 Sequence buffer full

No more steps can be added to the current sequence being edited.

Reduce the number of steps in the sequence.

Break up the testing into two or more separate sequences.