Degrees F.

A/D

Accuracy

Accuracy

Accuracy

 

 

 

 

 

 

At 1600 Degrees F with

TC Type

Min

Max

Full scale

Full Scale

At 1600 Degrees F.

thermister error

 

 

 

 

 

 

 

B

32

3308

0.0124%

0.0179%

0.29

0.65

 

 

 

 

 

 

 

C

32

4208

0.0062%

0.0067%

0.11

0.47

 

 

 

 

 

 

 

E

-328

1832

0.0031%

0.0032%

0.05

0.41

 

 

 

 

 

 

 

J

-346

2192

0.0031%

0.0036%

0.06

0.42

 

 

 

 

 

 

 

K

-328

2502

0.0031%

0.0045%

0.07

0.43

 

 

 

 

 

 

 

N

-328

2372

0.0031%

0.0052%

0.08

0.44

 

 

 

 

 

 

 

NNM

0

1409

0.0031%

0.0033%

0.05

0.41

 

 

 

 

 

 

 

R

-58

3214

0.0062%

0.0118%

0.19

0.66

 

 

 

 

 

 

 

S

-58

3214

0.0124%

0.0133%

0.21

0.57

 

 

 

 

 

 

 

T

-328

752

0.0062%

0.0119%

0.19

0.55

 

 

 

 

 

 

 

*** Note: Accuracy at 1600 Degrees F with thermister error requires SDS firmware level 1.10 or later ***

Software Installation

You will find the SDS Recorder software CD that ships with the unit includes all the reporting and administrative utilities. Insert the CD into the CD ROM device on your computer and click the setup.exe file. You will be given a step-by-step instruction for completing the installation process. See Installing the SDS Software section for more details.

Calibration

The SDS data logger is calibrated prior to shipment. The calibration process is available through menu option 22 on the data logger. Instructions for calibration are included in this manual. Super Systems calibrates the SDS data logger using NIST traceable instruments that are listed on the calibration certificate included with the device. The default calibration is performed using a zero and span routine for the millivolt range of 0 – 80mV. Each range that is being used should be calibrated. For a list of suggested mV ranges by thermocouple types, please refer to the section on Input Calibration in the manual. Other calibration ranges are available upon request.

An input device will be used to input a specific type K thermocouple value and the data logger value will be recorded.

This will be performed at different temperatures and printed with the certificate.