Super Systems Inc. Page 6 SDS Data Logger Manual #4565 Rev D
Degrees F. A/D Accuracy Accuracy Accuracy
TC Type Min Max Full scale Full Scale At 1600 Degrees F.
At 1600 Degrees F with
thermister error
B 32 3308 0.0124% 0.0179% 0.29 0.65
C 32 4208 0.0062% 0.0067% 0.11 0.47
E -328 1832 0.0031% 0.0032% 0.05 0.41
J -346 2192 0.0031% 0.0036% 0.06 0.42
K -328 2502 0.0031% 0.0045% 0.07 0.43
N -328 2372 0.0031% 0.0052% 0.08 0.44
NNM 0 1409 0.0031% 0.0033% 0.05 0.41
R -58 3214 0.0062% 0.0118% 0.19 0.66
S -58 3214 0.0124% 0.0133% 0.21 0.57
T -328 752 0.0062% 0.0119% 0.19 0.55
*** Note: Accuracy at 1600 Degrees F with thermister error requires SDS firmware level 1.10 or later ***
Software Installation
You will find the SDS Recorder software CD that ships with the unit includes all the reporting and
administrative utilities. Insert the CD into the CD ROM device on your computer and click the setup.exe
file. You will be given a step-by-step instruction for completing the installation process. See
Installing the SDS Software
section for more details.
Calibration
The SDS data logger is calibrated prior to shipment. The
calibration process is available through menu option 22 on the
data logger. Instructions for calibration are included in this
manual. Super Systems calibrates the SDS data logger using
NIST traceable instruments that are listed on the calibration
certificate included with the device. The default calibration is
performed using a zero and span routine for the millivolt range of
0 – 80mV. Each range that is being used should be calibrated.
For a list of suggested mV ranges by thermocouple types, please
refer to the section on Input Calibration in the manual. Other
calibration ranges are available upon request.
An input device will be used to input a specific type K
thermocouple value and the data logger value will be recorded.
This will be performed at different temperatures and printed with the certificate.