Chapter 4
PC-IP Operation
45
Figure
Limits are also used in the manual screens to provide Pass/Fail indicators on the graphic displays and to provide Pass/Fail criteria for numeric readouts. Limits are displayed as a yellow box on the graphic displays. Results within the yellow box are within limits. Results outside of the yellow box are outside of the limits and displayed as a red bar. Numeric readouts are displayed in red text when they are outside of the defined limits.
Database
The current database and path are displayed at the top of the screen. Once edited, the database may be saved by clicking on the database Save All button, or may be saved as a new database by using the Database/Save As feature from the tool bar.
Note: The database files are separate from the user’s tests result files. Saving one does not save the other. Be sure to use the Database menu for all database operations.
Channel Table
The Channel Table used for the tests at each location may be selected by clicking on the white box displaying the current channel table. The user may select from any of the channel tables listed in the