Chapter 4
PC-IP Operation
55
Leakage Limit - Select Set Llimit and use the arrow keys to increment the display to the desired limit to trigger the leakage alarm.
Distance - Select set and use the arrow keys to select the desired distance for the measurement (the range of values is 10 to 500 feet).
Compensation - Select Compensation and use the arrow keys to select the desired compensation value (+ or – 6.0 dB in 1 dB increments).
Signal Type - Select the type of signal being monitored and the type of tagging. Use the arrow keys to select CW or Analog moduaion with No, AM or FM Tagging.
Tag Frequency - This future enhancement will allow the user to select a TAG FREQUENCY and use the arrow keys to increment or the alphanumeric keys to enter the desired tag frequency from 5 to 2000 Hz.
Review the settings and click on the Database Save button to save the setup selections
Leakage Measurements
Leakage measurements will include the channel number, frequency, leakage measurement,