Calibration

Calibration Procedure

Calibration Procedure

Output Voltage

Calibration of programming and readback are combined in a single procedure.

1.Set the load to open circuit. Attach a DVM across the output terminals.

2.Minimum calibration level

Set the output voltage to 10% by sending the command:

CAL:OUTP:VOLT:LEV MIN

3.Enter voltage data

Enter the voltage read from the external DVM. CAL:OUTP:VOLT:DATA <voltage>

4.Maximum calibration level

Set the output voltage to 90% by sending the command:

CAL:OUTP:VOLT:LEV MAX

5.Enter voltage data

Enter the voltage read from the external DVM. CAL:OUTP:VOLT:DATA <voltage>

6.Power supply calculates and stores calibration constants.