Table 1.1 Specifications for HPD 300 W with GPIB-M or CANbus

Models

15-20

30-10

60-5

Program Resolution

 

 

 

Voltage

2.4mV

4.7mV

9.3mV

Current

2.8mA

1.4mA

0.7mA

 

 

 

 

Program Accuracy1

 

 

 

Voltage

60mV

70mV

90mV

 

±0.1%

±0.1%

±0.12%

Current

75mA

50mA

25mA

 

±0.12%

±0.12%

±0.1%

Readback Resolution

 

 

 

Voltage

2.4mV

4.7mV

9.3mV

Current

2.8mA

1.4mA

0.7mA

Readback Accuracy1

 

 

 

Voltage

45mV

90mV

175mV

 

±0.3%

±0.3%

±0.3%

Current

75mA

40mA

25mA

 

±0.12%

±0.12%

±0.1%

1.Apply accuracy specifications according to the following voltage program accuracy example: Set a model HPD 15-20 power supply to 10 volts.

The expected result will be within the range of 10 volts ± 60mV ± 0.1% of the set voltage of 10 volts.