Installation and Configuration
User Lines
User Lines
Connection
(XT, HPD,
XPD)
Figure 2.11User Signals Connector (XT, HPD and XPD)
Use a standard
Add a ferrite block to reduce radiated emission. The one inch square ferrite block with
To install the ferrite block:
1.Position the block no more than 5 cm (2 in.) from the power supply end of the user cable.
2.Open the ferrite block housing.
3.Loop the cable through the ferrite block. See Figure 2.12, “XT, HPD, XPD User Cable with Ferrite Block” on page 41.
4.Close the housing clip.
The ferrite block ensures that the power supply system meets radiated emission requirements for CE mark.
J7 User Cable
Ferrite Block
To J7 Connector | To User Custom Interface |
Figure 2.12XT, HPD, XPD User Cable with Ferrite Block
Release 2.1 | 41 |