Status Registers
Overview
Table 4.7 QUEStionable Status Register
Bit | Bit Weight | Bit Name | Description |
0 | 1 | VOLTage Summary | Reflects a summary of the |
|
|
| VOLTage |
1 | 2 | CURRent Summary | Reflects a summary of the |
|
|
| CURRent |
|
|
|
|
2 | 4 | TIME | Not implemented |
|
|
|
|
3 | 8 | Reserved |
|
|
|
|
|
4 | 16 | Over Temperature | Indicates the OTP circuit has |
|
|
| tripped. |
5 | 32 | FREQuency Summary | Not implemented |
|
|
|
|
6 | 64 | PHASe Summary | Not implemented |
|
|
|
|
7 | 128 | MODulation Summary | Not implemented |
|
|
|
|
8 | 256 | CALibration | Indicates an error in the unit |
|
|
| calibration. |
|
|
|
|
9 | 512 | Not Used | Not implemented |
|
|
|
|
10 | 1024 | Not Used | Not implemented |
|
|
|
|
11 | 2048 | AC Off | Indicates an AC Supply failure. |
|
|
|
|
12 | 4096 | UNRegulated | Indicates that the output is not |
|
|
| regulated in either Constant |
|
|
| Voltage mode, Constant Current |
|
|
| mode or Constant Power mode. |
|
|
| Reflects the inverse of the |
|
|
| Operation Regulation Summary bit. |
|
|
|
|
13 | 8192 | INSTrument Summary | Not implemented |
|
|
|
|
14 | 16384 | Command Warning | Not implemented |
|
|
|
|
15 | 32768 | Not Used | Always zero |
|
|
|
|
VOLTage This shows whether the present voltage level is over or under the specified trip limit.
| VOLTage |
| ||
| Bit | Bit Weight | Bit Name | Description |
0 | 1 | Over VOLtage | Set if the supply’s output voltage | |
|
|
|
| exceeds the |
|
|
|
| either |
|
|
|
| or the fixed trip limit. |
1 | 2 | Under VOLtage | Set if the supply’s output voltage is less | |
|
|
|
| than the |
|
|
|
| trip level (variable trip limit) and the |
|
|
|
| supply is in Operation state. |
76 | Operating Manual for Multichannel Functionality |