Configure
Output
Protection
Operation
Power Supply Operation
Five configurable protection mechanisms are available:
•OVP:
•UVP:
•OCP:
•UCP:
•Fold: Fold Protection. (See below.)
OVP shuts down the power supply if the protection limit is exceeded. The other options offer a choice: they shut down the power supply or issue a warning by setting a bit in the status register. Set the protection mechanism to ON to enable it to shutdown the power supply.
When the protection level is set to zero, that mechanism is disabled. However, in the case of OVP, a fixed hardware protection mechanism still exists to protect the supply from damage.
To set the
SOUR:VOLT:PROT <voltage>
To set the
SOUR:VOLT:PROT:UND <voltage>
To enable UVP shutdown:
SOUR:VOLT:PROT:UND:STAT
To set the
SOUR:CURR:PROT <current>
To enable OCP shutdown:
SOUR:CURR:PROT:STAT
To set the
SOUR:CURR:PROT:UND <current>
To enable UCP shutdown:
SOUR:CURR:PROT:UND:STAT
To check if a protection mechanism was tripped:
SOUR:VOLT:PROT:OVER:TRIP?
SOUR:VOLT:PROT:UND:TRIP?
SOUR:CURR:PROT:OVER:TRIP?
SOUR:CURR:PROT:UND:TRIP?
Alternatively, you can query the status register. See Appendix B. “SCPI Command Reference”.