Installation and Configuration
Configure for GPIB Operation
Configure for GPIB Operation
Change
Remote
Control
Source
The power supply is ready for GPIB communication when it leaves the factory, but you may change any of the control settings. The default remote control setting is GPIB, address 2.
Select an interface for remote control.
SYST:REM:SOUR [GPIBMCHannel]
where
•GPIB: control via GPIB interface
•MCHannel: multichannel operation using the CANbus interface
Set GPIB Change the GPIB address.
Address SYST:COMM:GPIB:ADDR
where
Be sure to make a note of the GPIB address. If the address is unknown, use a feature such as “Scan for Instruments” on National Instruments’ Measurement and Automation Explorer.
34 | Operating Manual for Multichannel Functionality |