Operation
Power Supply Operation
User Settings If you have a frequent or constant need for a specific voltage and current output, you (Save and can save these setpoints in the power supply’s memory as a user setting. Once a
Recall) setting is stored, it remains in the power supply’s memory after the unit is powered off.
Ten user setting memory locations are available, and each saves the following parameters:
•Voltage setpoint
•Current setpoint
•
•
•
•
•Foldback protection mode and delay settings
•AC Off auto recovery configuration (ACO)
•OTP
•Aux line configuration
•Voltage and current soft limits
•Triggered setpoints and trigger source
Note All parameters are saved and used when the user setting is recalled. Therefore, you should set parameters that you do not care about to the factory defaults.
To store (save) settings:
*SAV
or the SCPI equivalent for multichannel addressed units
SYST<channel>:SAVE
where
Example:
*SAV 1 will save all the present settings to user setting location 1.
To recall settings:
*RCL
To save default settings:
*SDS
Release 2.1 | 55 |