Operation

Auto Sequencing

Set V, I, and P The voltage, current and power setpoints can be limited to less than the supply rating Limits range to match the tolerance of connected equipment or any other criteria you may

have.

Once the limits have been changed from the supply’s default rated output, settings outside this range are no longer accepted.

Note Setpoint limits do not apply to the triggered outputs and auto sequence outputs.

To set the voltage limits:

SOUR:VOLT:LIM:HIGH <voltage>

SOUR:VOLT:LIM:LOW <voltage>

To set the current limits:

SOUR:CURR:LIM:HIGH <current>

SOUR:CURR:LIM:LOW <current>

If the high end of the range was set to 5 volts, the command, SOUR:VOLT 10 would return an error.

Triggered To set a triggered setpoint:

Setpoints SOUR:VOLT:TRIG <voltage>

SOUR:CURR:TRIG <current>

See “Triggering Commands” on page 66 for more information.

Release 2.1

65